With the in-depth development of the infrared physics and infrared technology
the measurement of target radiation characteristics is widely appreciated. And the infrared detector radiometric calibration is a fundamental work. This paper establish a radiometric calibration system
the absolute radiometric calibration of the 3~5 m infrared focal plane arrays. Firstly
give the radiometric calibration mathematical model
developed a calibration process and the radiometric calibration experiment. As infrared standard source and the impact of environmental volatility
a weighted least squares radiation calibration optimization is proposed. The radiation characteristic measurement experiment was carried out outside the field. Results show that the inversion error of optimization method is better than 4%. The accuracy of radiometric calibration was improved. And the results meet the actual needs of the project.
关键词
Keywords
references
Jacobs P A. Thermal Infrared Characterization of Ground Targets and Backgrounds [M]. Washington:SPIE,1996.[2] Wack E C. Pre-launch infrared of the GOES I-M Imager and Sounder [J].SPIE,2004,4135:140-149.[3] Kiemme B J. A 1 cm collimated source for use in Infrared calibrations [J]. SPIE,2000,4028:411-419.[4] 王锐,宋克非.高精度紫外探测器辐射定标系统 [J]. 光学 精密工程, 2009,17(3):469-474.[5] 马宏,王金波. 误差理论与仪器精度 [M].北京:兵器工业出版社,2007.[6] 李步蟾,肖峻.基于定标的红外焦平面非均匀性校正算法研究 [J]. 红外, 2008,25(6):5-9.[7] 李宪圣,任建伟,张立国,等.大口径红外光电系统现场辐射定标装置的研制 [J]. 光电子·激光, 2006,17(2):175-178.[8] 林晓钢,顾乃庭,杨泽平.基于小孔夫琅和费衍射法的CCD光电响应特性标定研究 [J]. 光学 精密工程, 2008,16(3):410-414.[9] 陈风,郑小兵.光谱非匹配对于光学遥感器定标精度影响的分析 [J]. 光学 精密工程, 2008,16(3):415-419.[10] 赵玉环,闫丰,周跃,等.紫外ICCD的辐射定标 [J]. 光学 精密工程, 2008,16(9):1572-1576.